Development of a compact E×B microchannel plate detector for beam imaging
No Thumbnail Available
Can’t use the file because of accessibility barriers? Contact us
Date
2017-09-18
Journal Title
Journal ISSN
Volume Title
Publisher
Permanent Link
Abstract
A beam imaging detector was developed by coupling a multi-strip anode with delay line readout to an E×B microchannel plate (MCP) detector. This detector is capable of measuring the incident position of the beam particles in one-dimension. To assess the spatial resolution, the detector was illuminated by an α-source with an intervening mask that consists of a series of precisely-machined slits. The measured spatial resolution was 520μm FWHM, which was improved to 413μm FWHM by performing an FFT of the signals, rejecting spurious signals on the delay line, and requiring a minimum signal amplitude. This measured spatial resolution of 413μm FWHM corresponds to an intrinsic resolution of 334μm FWHM when the effect of the finite slit width is de-convoluted. To understand the measured resolution, the performance of the detector is simulated with the ion-trajectory code SIMION.
Description
Keywords
Citation
Wiggins, Blake, et al. "Development of a compact E×B microchannel plate detector for beam imaging." Nuclear Instruments and Methods in Physics Research Section A, vol. 866, pp. 202-206, 2017-9-18, https://doi.org/10.1016/j.nima.2017.06.006.
Journal
Nuclear Instruments and Methods in Physics Research Section A