Enhanced retinal vasculature imaging with a rapidly configurable aperture
No Thumbnail Available
Can’t use the file because of accessibility barriers? Contact us
Date
2018-02-23
Journal Title
Journal ISSN
Volume Title
Publisher
Permanent Link
Abstract
In adaptive optics scanning laser ophthalmoscope (AOSLO) systems, capturing multiply scattered light can increase the contrast of the retinal microvasculature structure, cone inner segments, and retinal ganglion cells. Current systems generally use either a split detector or offset aperture approach to collect this light. We tested the ability of a spatial light modulator (SLM) as a rapidly configurable aperture to use more complex shapes to enhance the contrast of retinal structure. Particularly, we varied the orientation of a split detector aperture and explored the use of a more complex shape, the half annulus, to enhance the contrast of the retinal vasculature. We used the new approach to investigate the influence of scattering distance and orientation on vascular imaging.
Description
Keywords
Citation
Sapoznik, K. A., et al. "Enhanced retinal vasculature imaging with a rapidly configurable aperture." Biomedical optics express, vol. 9, no. 3, pp. 1323-1333, 2018-02-23, https://doi.org/10.1364/BOE.9.001323.
Journal
Biomedical optics express